Figure 1
Which edition of the audiometer standard each federal rule names
| OSHA, 29 CFR 1910.95(h)(2) | ANSI S3.6-1969 | Workplace audiometers must meet this edition and be maintained and used in line with it |
|---|---|---|
| FDA regulation, 21 CFR 874.1050(b) | ANSI S3.6-1996 | Class II audiometer exempt from premarket notification if it complies; otoacoustic emission devices excluded |
| FDA recognized standard 4-275 | ANSI ASA S3.6-2018 (Reaffirmed 2023) | Declarations of conformity accepted until December 19, 2027 |
| FDA recognized standard 4-350 | ANSI ASA S3.6-2025 | Recognized in full, entered December 22, 2025; supersedes recognition 4-275 |
- FDA,accessdata.fda.gov/…/detail.cfm
- FDA,accessdata.fda.gov/…/detail.cfm
- eCFR (Office of the Federal Register),ecfr.gov/…/section-874.1050 (official API)
- OSHA,osha.gov/…/1910.95
Figure 2
OSHA calibration schedule for workplace audiometers
| Functional check | Before each day's use | Deviations of 10 decibels or greater require an acoustic calibration |
|---|---|---|
| Acoustic calibration check (appendix E) | At least annually | Deviations of 15 decibels or greater require an exhaustive calibration |
| Exhaustive calibration (S3.6-1969 sections) | At least every two years | Scheduled, whatever the earlier results |
- OSHA,osha.gov/…/1910.95
How to read these numbers (2)
- Each row restates the rule's own text; the standard's numerical content is not reproduced.
- The OSHA schedule applies to employers' audiometric testing programs, not to clinics in general.
What the FDA calls a hearing test device
In FDA's device classification database, product code EWO carries the device name "Audiometer," regulation number 874.1050 and device class 2. The regulation, 21 CFR 874.1050, defines the device this way: "An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals." Both the manual and the automated machine fall under the same rule.
The second paragraph of the rule sets the class and a condition. The audiometer is Class II, and "except for the otoacoustic emission device, the device is exempt from the premarket notification procedures in subpart E of part 807" "if it is in compliance with American National Standard Institute S3.6-1996," within limits the rule lists. Put simply, an audiometer that complies with that edition doesn't need a premarket notification, while the otoacoustic emission device, the one used for an OAE hearing test, isn't covered by that exemption.
Which edition of ANSI S3.6 each rule names
ANSI S3.6 is the American National Standard "Specification for Audiometers." Federal rules don't all cite the same edition, and the table above lines them up. OSHA's noise rule still names S3.6-1969 for workplace audiometers. The FDA regulation's exemption names S3.6-1996. FDA's list of recognized consensus standards, which makers use to declare conformity in premarket submissions, now carries ANSI ASA S3.6-2025 as recognition 4-350, with a date of entry of December 22, 2025.
The change of edition has a deadline. FDA writes that it "will accept declarations of conformity, in support of premarket submissions, to [Rec 4-275] until December 19, 2027," recognition 4-275 being the 2018 edition reaffirmed in 2023. After that date, a declaration to the 2018 edition "will not be accepted." The deadline concerns makers' submissions to FDA, not audiometers already in use.
FDA's summary of the 2018 edition also explains the types printed on spec sheets: "Audiometers used for diagnostic assessment are classified as Types 1, 2, or 3, while audiometers having air-conduction facilities only are classified as Type 4."
How often a workplace audiometer must be calibrated
Employers who run audiometric testing under OSHA's occupational noise rule, 29 CFR 1910.95, follow a written calibration schedule in paragraph (h)(5). It has three steps, shown in the second table.
Every day, "the functional operation of the audiometer shall be checked before each day's use by testing a person with known, stable hearing thresholds," and by listening for distorted or unwanted sounds. If that check is off by 10 decibels or more, the audiometer goes to an acoustic calibration. At least once a year, calibration is checked acoustically under the rule's appendix E, and a deviation of 15 decibels or more calls for an exhaustive calibration. Whatever those checks show, an exhaustive calibration is due at least every two years, against listed sections of S3.6-1969.
The same rule says what the machine has to test: pure tone, air conduction thresholds, at 500, 1000, 2000, 3000, 4000 and 6000 Hz at a minimum, each ear separately. Noise limits that trigger the program are on our safe decibel levels page.
Who runs a workplace hearing test, and in what room
The same OSHA rule sets conditions around the device. Workplace audiograms must be done by "a licensed or certified audiologist, otolaryngologist, or other physician, or by a technician who is certified by the Council of Accreditation in Occupational Hearing Conservation," or by a technician who has shown competence, including in checking calibration.
Two kinds of machine carry an extra requirement: "pulsed-tone and self-recording audiometers, if used, shall meet the requirements specified in appendix C." The room matters too, since examinations "shall be administered in a room meeting the requirements listed in appendix D." And each employee's audiogram record has to show the "date of the last acoustic or exhaustive calibration of the audiometer," so the device's calibration history travels with every test result.
The hearing tests an audiometer runs, by billing code
Billing codes split audiometry by what is tested. In the CMS 2026 physician fee schedule file, 92551 is labelled "Pure tone hearing test air," 92552 "Pure tone audiometry air," 92553 "Audiometry air & bone," 92555 "Speech threshold audiometry," 92556 "Speech audiometry complete" and 92557 "Comprehensive hearing test." These are CMS short labels, not the full CPT descriptions.
Each code pays for the test, not the machine. What Medicare allows for each one is on our hearing test cost page, and whether Medicare covers the visit is on does Medicare cover a hearing test.
Frequently asked questions
What device is used to test hearing?
An audiometer. FDA's classification database names the device "Audiometer" under product code EWO, and the regulation covers both the audiometer and the automated audiometer. An otoacoustic emission device sits in the same regulation but is left out of its premarket exemption.
What is the instrument used to measure hearing called?
It's called an audiometer, the word used in the title of its standard, ANSI ASA S3.6-2025, "Specification for Audiometers." FDA's summary of the 2018 edition sorts audiometers into Types 1, 2 or 3 for diagnostic assessment and Type 4 for air-conduction-only machines.
What are the three types of hearing tests?
Billing codes give one way to split them: pure tone air conduction only (92552), air and bone conduction (92553), and speech audiometry (92556 for the complete test). They are CMS labels for billing, not a clinical ranking of tests.
Sources
Read on November 3, 2026: FDA's classification record for product code EWO, the eCFR text of 21 CFR 874.1050, FDA's recognized-standard records for ANSI ASA S3.6-2025 and S3.6-2018, OSHA's 29 CFR 1910.95, and the CMS 2026 physician fee schedule file. We don't test or rate equipment.
- https://api.fda.gov/device/classification.json?search=product_code:EWO&limit=1
- https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-874/subpart-B/section-874.1050 (official API)
- https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfstandards/detail.cfm?standard__identification_no=46554
- https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfStandards/detail.cfm?standard__identification_no=46555
- https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.95
- https://www.cms.gov/files/zip/rvu26d-updated-08-26-2026.zip
Cost and coverage research, not medical advice. This page describes federal device and workplace rules; which hearing test a person needs is for a hearing care professional.